Stat Fax 4700 Service Manual =link= -
If the tungsten-xenon lamp is replaced, it must be perfectly aligned with the guide bracket aperture before refastening the bracket screws.
Accessing internal components requires inverting the instrument and removing the enclosure screws using a Philips #1 screwdriver.
Features a linear measurement range of 0.0 to 3.0 Absorbance (A). stat fax 4700 service manual
A non-volatile memory stores up to 120 user-programmed tests.
is a compact, microprocessor-controlled EIA microstrip reader designed for modern laboratories. This standalone system is ideal for routine ELISA assays, offering bichromatic optics, an interactive touchscreen, and an onboard printer for rapid results. If the tungsten-xenon lamp is replaced, it must
Standard filters include 405, 450, 492, and 630 nm, with options for up to six filters ranging from 340 nm to 700 nm.
It typically reads, calculates, and prints results for a 12-well strip in approximately 30 seconds. Service & Maintenance Procedures A non-volatile memory stores up to 120 user-programmed tests
The Auto-Track™ transport system can load up to three strips (up to 36 wells total) and handles break-apart, single, or double strips.